User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2517 Verification Proceedure (Ver1.0) DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 80 CONFIGURATION: Fluke 5700A ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM P/N: @ * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Auto Calibration * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P M F L W 1.002 HEAD { INITIAL PROCESS } 1.003 DISP CONNECT THE PM 2517X TO THE 5700 1.003 DISP BEFORE PERFORMING THE VERIFICATION 1.003 DISP AND ALLOW THE UUT TO WARM UP A FEW MINUTES 1.004 HEAD { ZERO TEST } 1.005 DISP SET THE UUT TO AUTORANGE VDC RANGE 1.005 DISP [27][91]1m CONNECT AS FOLLOWS 1.005 DISP ! ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 1.005 DISP ! [27][91]1m 5700 UUT 1.005 DISP ! SENSE HI ÄÄÄÄÄÄÄÄÄ¿ 1.005 DISP ! OUTPUT HI ÄÄÄÄÄÄÄÄÄÁÄ> VOLT/OHM INPUT 1.005 DISP ! OUTPUT LO ÄÄÄÄÄÄÄÄÄÂÄ> COMMON 1.005 DISP ! SENSE LO ÄÄÄÄÄÄÄÄÄÙ 1.006 5700 1 0mV 0.3U 2W 2.001 HEAD { DC VOLTAGE PERFORMANCE VERIFICATION} 2.002 5700 1 900mV 0.34% 2W 3.001 5700 10 9V 0.34% 2W 4.001 5700 100 90V 0.34% 2W 5.001 5700 1000 900V 0.34% 2W 6.001 HEAD 6.002 DISP SELECT ALTERNATING VOLTAGE 6.003 HEAD {ALTERNATING VOLTAGE PERFORMANCE VERIFICATION} 6.004 5700 1 900mV 0.7% 60H 2W 7.001 5700 10 9V 0.7% 60H 2W 8.001 5700 10 9V 1.2% 5kH 2W 9.001 5700 100 90V 0.7% 60H 2W 10.001 5700 1000 600V 1.00% 60H 2W 11.001 HEAD {OHMS PERFORMANCE VERIFICATION} 11.002 DISP SELECT THE OHMS FUNCTION 11.003 5700 100Z 1.6% 2W 12.001 5700 1kZ 1.6% 2W 13.001 5700 10kZ 1.6% 2W 14.001 5700 100kZ 2.1% 2W 15.001 5700 1MZ 2.2% 2W 16.001 HEAD {LOW CURRENT PERFORMANCE VERIFICATION} 16.002 DISP SELECT THE DC CURRENT FUNTION 16.003 5700 90mA 0.62% 2W 17.001 DISP SELECT THE AC CURRENT FUNCTION 17.002 5700 90mA 0.92% 60H 2W 18.001 HEAD {HIGH CURRENT PERFORMANCE VERIFICATION} 18.002 DISP [7][D500][7][D500][7][D500][27][91]1m REWIRE INSTRUCTION 18.002 DISP REWIRE THE UUT FROM THE V-mA INPUT TO THE 10A INPUT 18.003 DISP [27][91]1mDID YOU REWIRE, PLEASE DON'T BLOW THE FUSE 18.004 5700 1.9A 0.9% 60H 2W 19.001 DISP SELECT THE DIRECT CURRENT FUNCTION 19.002 5700 1.9A 1.1% 2W 20.001 DISP [27][91]1m SELECT VOLTAGE DC FUNCTION 20.002 END